New! Sign up for our free email newsletter.
Science News
from research organizations

Reducing Patient Radiation Dose During Pulmonary CT Angiography

Date:
June 30, 2009
Source:
American Roentgen Ray Society
Summary:
While screening for possible pulmonary emboli using pulmonary CT angiography, a new study shows that radiologists can effectively lower the patient radiation dose by approximately 44 percent, and improve vascular enhancement without deterioration of image quality, according to a new study.
Share:
FULL STORY

While screening for possible pulmonary emboli using pulmonary CT angiography, a new study shows that radiologists can effectively lower the patient radiation dose by approximately 44% and improve vascular enhancement without deterioration of image quality, according to a study performed at Brigham and Women’s Hospital and Harvard Medical School, Boston, MA.

A total of 400 patients believed to have a pulmonary embolism were evaluated using pulmonary CT angiography. Two hundred patients were evaluated using the standard peak kilovoltage setting of 130 or 120 kVp and the other 200 patients were evaluated using a low peak kilovoltage setting of 110 or 100 kVp. “Results showed that lowering the peak kilovoltage setting by 20-kVp lead to superior vascular enhancement without deterioration of image quality—allowing us to effectively reduce the patient radiation dose,” said Shin Matsuoka, MD, lead author of the study.

“CT has become an essential tool for the diagnosis of pulmonary embolism. However because of the high percentage of negative results, radiation exposure has become an important issue. Our study shows that lowering the kilovoltage setting may be an effective method of lowering the radiation dose for most patients,” he said.

“Lowering the kilovoltage setting is something that could be easily incorporated into daily clinical practice because there is no additional equipment needed and there are no extra costs,” said Dr. Matsuoka.


Story Source:

Materials provided by American Roentgen Ray Society. Note: Content may be edited for style and length.


Journal Reference:

  1. Matsuoka et al. Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings. American Journal of Roentgenology, 2009; 192 (6): 1651 DOI: 10.2214/AJR.08.1730

Cite This Page:

American Roentgen Ray Society. "Reducing Patient Radiation Dose During Pulmonary CT Angiography." ScienceDaily. ScienceDaily, 30 June 2009. <www.sciencedaily.com/releases/2009/06/090630143525.htm>.
American Roentgen Ray Society. (2009, June 30). Reducing Patient Radiation Dose During Pulmonary CT Angiography. ScienceDaily. Retrieved December 21, 2024 from www.sciencedaily.com/releases/2009/06/090630143525.htm
American Roentgen Ray Society. "Reducing Patient Radiation Dose During Pulmonary CT Angiography." ScienceDaily. www.sciencedaily.com/releases/2009/06/090630143525.htm (accessed December 21, 2024).

Explore More

from ScienceDaily

RELATED STORIES